Menu structure – Det-Tronics UD10 FlexVu Explosion-Proof Universal Display Unit User Manual

Page 57

Advertising
background image

4.2

K-5

95-8661

Calibration procedure
To calibrate the NTMOS detector with the FlexVu UD10

Display:

1. Touch the magnet to the ENTER/SELECT button to

display the Main Menu. Follow the illustration below

to navigate to the “Calibrate” menu.

Main Menu
Process Vars
Display Status
Device Status
Display Setup
Device Setup
Device Cal
Display Test

Device Cal

Cal Gas Conc
Calibrate

Calibrate

Execute
Abort

2. Activate “Execute” (Enter/Select) to begin the zero

calibration.

3. The UD10 will display “Waiting for Zero” on the main

display screen.

4. When zero calibration is complete (approximately

one minute), the UD10 will display “Waiting for

Span” on the main display screen.

5. Apply calibration gas to the detector.

6. With 50 ppm H

2

S applied to the detector, the UD10

display will continue to show “Waiting for Span” while

the span calibration is being performed.

7. When the UD10 Display shows “Remove Cal Gas”

the calibration is complete. Remove calibration gas

from the detector.

8. When the gas level falls below the lowest alarm

setpoint, the UD10 automatically exits the Calibrate

mode and returns to normal operating mode.

menu structure

UD10 with NTmOS H

2

S Detector

Refer to the following menu when using the UD10’s

LCD display and internal magnetic switches.

When connecting a HART Communicator to the UD10’s

4-20 mA output, refer to the “UD10 HART” menu in

Appendix E.

menu HeLp

Status menus only allow the user to view the data.
The Setup menus allow the user to both view and
edit the data.

Humidification Tube Calibration Kit Attached to nTMoS Detector

UD10 mA Output During Calibration

(UD10 with NTmOS)

UD10 Display Reading

Standard

mode

Replicate

mode

Waiting for Zero

3.8

2.2

Waiting for Gas

3.8

3.8

Waiting for Span

3.8

3.8

Remove Cal Gas

3.8

3.8

back to normal

4.0

4.0