3 installation with 2 3/4" feedthrough – INFICON Sputtering Sensor User Manual

Page 38

Advertising
background image

3 - 2

IP

N 07

4-

15

7L

Sputtering Crystal Sensor Operating Manual

CAUTION

Applicable to sensor 750-446 only: Use of more than 80"
(2 meters) of 1/8" tubing between the valve and the
bellows may cause switcher failure because of the time
required to bleed out sufficient air during
de-pressurization. If tubing greater than 1/8" is used, the
maximum length must be reduced proportional to the
additional volume.

8

Attach the valve’s intake (normally closed (NC) port) to the 80-90 PSIG (5.5-6.2
bar) [550-620 kPa] source of air. Or, 90-95 PSIG (6.2-6.55 bar) [620-655 kPa]
air supply for the Sputtering Shutter (IPN 750-005).

3.3 Installation with 2 3/4" Feedthrough

If you are installing the solenoid valve assembly with a dual sensor, a
2 3/4" feedthrough inclusive of two coaxial feedthroughs (IPN 002-080, see

Figure

5-4 on page 5-5

) is required. All other shuttered sensors utilizing 2 3/4"

feedthroughs require only a single coaxial feedthrough. The second coaxial
feedthrough is not used, and should be protected from damage as a result of
process material. Follow the steps below:

1

Install the Feedthrough.

2

Add the valve bracket (modified) to the desired location utilizing two of the
flange’s 1/4" clamp bolts.

3

Tighten the flange bolts.

4

Install the air fitting to the female thread adapter.

5

Connect the 1/8" air tube from the valve outlet to the feedthrough fitting. See
the caution at the top of this page.

6

Attach the valve’s intake (normally closed (NC) port) to the 80-90 PSIG (5.5-6.2
bar) [550-620 kPa] source of air. Or, 90-95 PSIG (6.2-6.55 bar) [620-655 kPa]
air supply for the Sputtering Shutter (IPN 750-005).

NOTE: Maximum temperature for the shutter control valve assembly is 105 °C for

bakeout and operation.