5 sam-242 card connections, Caution – INFICON SQM-242 Thin Film Deposition Controller Card Operating Manual User Manual
Page 27
2 - 5
IP
N 07
4-
54
9-
P1
A
SQM-242 Operating Manual
2.5 SAM-242 Card Connections
The input and output and connectors on the SAM-242 card are identical to those
on the SQM-242 card.
Figure 2-5 SAM-242 Card
Input Connections
BNC cables connect the SAM-242 input to the signals to be measured. The
SAM-242 accepts input voltages within +/- 10 V (dc).
NOTE: You can not connect sensors to these inputs.
CAUTION
The BNC connector shield of each SAM-242 input is
connected to a common analog ground. Input signals to
the SAM-242 must be within +/-10 V (dc) and share a
common ground. Failure to observe this constraint can
result in damage to your equipment and/or the SAM-242
card,
Output Connection
The SAM-242 outputs are identical to the SQM-242. See the previous section for
hookup instructions.