Monitor . . . . . . . . . . . . . . . . . . . . . . Monitor sensors halt deposition when their

Thickness setpoint is reached.

Figure 3-7 Sensor Setup

Often sensors are configured to tightly control the deposition rate of a material,
such as Sensor 1 & 2 above. However, you might also use a monitor sensor near
the substrate, to more accurately monitor the final thickness of the co-deposited
material.

Setpoint . . . . . . . . . . . . . . . . . . . . . . The material thickness (in kÅ) measured by a

monitor sensor that will halt deposition.

Analog . . . . . . . . . . . . . . . . . . . . . . . The SAM-242 analog input card measures

DC voltages in the +/-10 volt range. These
voltages may represent temperature, flow, or
any other process variable. The analog
frame allows you to modify the display to
show values in the desired units, using a
linear (y = mx + b) transformation.

Assume you have a temperature transmitter
that sends 0 V at 0°C and 10 V at 100°C. You
want to control temperature to 200°F (it’s an
example!). Set the analog input Gain to 18,
Offset to 32, and Units to Deg F (F = 9/5C +
32). The SQM-242 will display setpoints and
measurements associated with the analog
input in degrees F.

To leave the analog input display in Volts, set
Gain = 1 and Offset = 0.

Sensor 1 Controls

Material 1 Rate

Sensor 2 Controls

Material 2 Rate

Sensor 3 Monitors
CoDeposited Material