SQM-242 Operating Manual

7.2.4 Sputtering Sensor

Observe the general precautions for replacing crystals and follow the instructions
below to replace the crystal in a sputtering sensor.

1

Grip the body assembly with your fingers and pull it straight out to separate it
from the water-cooled front cover. (You may have to disconnect the sensor
cable in order to separate the parts.) See

Figure 7-5

.

2

Pull the crystal holder straight out from the front of the body assembly.

3

Remove the ceramic retainer from the crystal holder by pulling it straight out
with the crystal snatcher (see

section 7.2.5 on page 7-15

).

4

Turn the crystal holder over so that the crystal drops out.

5

Install a new crystal into the crystal holder with the patterned electrode facing
the back and contacting the leaf springs on the ceramic retainer.

6

Put the ceramic retainer back into the crystal holder and put the holder into the
body assembly of the sensor.

7

Align the position of the body assembly so that the connector matches with the
notch on the front cover of the sensor. Snap the two parts together. Reconnect
the sensor cable if it has been disconnected.

Figure 7-5 Sputtering Crystal Sensor

Body Assembly

IPN 750-619-G1

In-Vacuum Cable Assembly

(29 in.) IPN 007-044

Ceramic Retainer

IPN 007-023

Crystal

IPN 008-009-G10

(Silver)

Crystal Holder

IPN 007-049

Sensor

Front Cover

IPN 007-047