Section, Section 7.3.1.3 on – INFICON SQM-242 Thin Film Deposition Controller Card Operating Manual User Manual

Page 73

Advertising
background image

7 - 19

IP

N 07

4-

54

9-

P1

A

SQM-242 Operating Manual

11

Connect the six-inch BNC cable to the Feed-Through and disconnect it from the
oscillator.

12

Verify electrical continuity from the center pin of the Microdot connector on the
Feed-Through to the un-terminated end of the six-inch BNC cable.

13

Verify electrical isolation from the center pin to electrical ground (Feed-Through
body).

If the Feed-Through/six-inch BNC cable system is found to be defective, look for
defective contacts at the Feed-Through to BNC cable connection. Repair or
replace the Feed-Through as necessary, re-attach the BNC cable to the XIU and
In-Vacuum cable to the Crystal head and repeat this procedure starting at step 2.

7.3.1.3 Sensor Head Or Monitor Crystal

Diagnostic Procedure

NOTE: The procedure is for use with front load style sensor heads.

1

Remove the Crystal Cover from the Sensor Head.

2

Refer to

Figure 7-7 on page 7-16

. Connect the Crystal Sensor Emulator to the

Sensor Head at Point C. Note that this only works on a Front Load style sensor
head.

If the Crystal Fail message disappears after approximately 5 seconds the
Sensor Head is operating properly. Re-insert the Crystal Cover into the
Sensor Head.

If the Crystal Fail message remains, continue at step 3.

3

Disconnect the In-Vacuum cable from the Sensor Head and the Feed-Through.
Remove the Crystal Sensor Emulator from the Sensor Head.

4

Using an Ohm meter, verify the electrical connections on the Sensor Head.

Verify there is electrical continuity from the center pin contact on the
Microdot connector on the Sensor Head to the leaf spring contact in the
Sensor Head. Take care not to apply to much pressure on the center pin of
the microdot connector as it may become damaged.

There must be electrical isolation between the center pin of the Microdot
connector and the Sensor Head body.

If the Sensor Head is found to be defective, contact INFICON to have the Sensor
Head repaired.

Advertising