3 view menu: card setup – INFICON SQM-242 Thin Film Deposition Controller Card Operating Manual User Manual

Page 38

Advertising
background image

3 - 10

IP

N 07

4-

54

9-

P1

A

SQM-242 Operating Manual

Gain . . . . . . . . . . . . . . . . . . . . . . . . . The gain term for transforming voltage to

measured units. This is the m term in
y = mx + b.

Offset. . . . . . . . . . . . . . . . . . . . . . . . The offset term for transforming voltage to

measured units. This is the b term in
y = mx + b.

Units . . . . . . . . . . . . . . . . . . . . . . . . The units that you wish to display for the

analog input.

Monitor . . . . . . . . . . . . . . . . . . . . . . An analog input can also act as a monitor to

stop deposition. For example, an analog
signal from an optical monitor could stop
deposition when a certain voltage is reached.
A voltage input from a pressure transducer
might also prevent deposition until a certain
vacuum is reached.

Setpoint. . . . . . . . . . . . . . . . . . . . . . The voltage measured by a monitor input that

will halt deposition. Analog setpoints are
entered in Volts, not calculated units!

3.6.3 View Menu: Card Setup

This dialog controls the most basic functions of the SQM-242 card. It also provides
useful installation and troubleshooting information.

Simulate . . . . . . . . . . . . . . . . . . . . . Normally the SQM-242 card uses the quartz

crystals as inputs for controlling the source
outputs. The SAM-242 card uses analog
input voltages for control. Simulate mode
simulates these inputs. No SQM-242 or
SAM-242 card needs to be installed for the
simulate mode.

Card 1, Card 2, Analog. . . . . . . . . . Shows the firmware revision of each installed

card. A value of 0.00 indicates that the card
is not seen by the software, and is probably
not installed properly in Windows. See

Chapter 7, Troubleshooting and
Maintenance

.

Advertising