A tot, Table 5-23 on, For more in – INFICON Composer Elite Gas Concentration Monitor User Manual

Page 100: Lue see, Table 5-23, Urrent warning value refer to, Refer to, Table

Advertising
background image

5 - 34

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

12

Heater 2 Current -
Expectations Not Met

0x0000 0800

SC

SS

Self Clear ACTION(SC): SC shuts off

Heater.
ACTION(CC): CC may
reallocate any power
management resources
assigned to the SC.
NOTE: Does not meet
expectation based on PWM duty

11

Heater 1 Setpoint -
Bad Parameter

0x0000 0400

SC

SS

Cleared on

read

NOTE: Attempt made to set
parameter above operating
range (>65°C).

10

Heater 2 Setpoint -
Bad Parameter

0x0000 0200

SC

SS

Cleared on

read

NOTE: Attempt made to set
parameter above operating
range (>68°C).

Table 5-23 System and hardware warnings

Warning

Number

Description

Warning Code Set by

CC

1

or SC

2

Sensor

Specific

(SS

3

)

Set/Clear

Trigger

General Notes/Actions

32

Composer Elite
Temperature Outside
Operating Range

0x8000 0000

CC

Self clear

31

DeviceNet Installed
but Not Operating

0x4000 0000

CC

Self clear NOTE: Indicates that the

DeviceNet module was found
but DeviceNet system is not
currently operating.

30

DeviceNet Failed to
Transmit a Response

0x2000 0000

CC

Cleared on

read

NOTE: Failed to transfer
response from CC to ABIC

4

module.

29

Parameters Reset

0x1000 0000

CC

Cleared on

read

NOTE: Could be result of
automatic parameter reset due
to compatibility error or from a
manual reset.

28

CC Firmware
Changed

0x0800 0000

CC

Cleared on

read

27

Splash Image
Incompatible or Invalid

0x0400 0000

CC

Cleared on

read

NOTE: The splash image read is
not compatible with the system
expectations and thus cannot be
displayed.

26

Parameters Upgraded 0x0200 0000

CC

Cleared on

read

1

CC = Controller Card,

2

SC = Sensor Card,

3

SS = Specific Sensor,

4

ABIC = Anybus Integrated Circuit,

5

R2= Lorentzian Coefficient of Determination

Table 5-22 System and hardware errors (continued)

Error

Number

Description

Error Code

Set by

CC1

or SC2

Sensor

Specific

(SS3)

Set/Clear

Trigger

General Notes/Actions

1

CC = Controller Card,

2

SC = Sensor Card,

3

SS = Specific Sensor

Advertising