INFICON Composer Elite Gas Concentration Monitor User Manual

Page 75

Advertising
background image

5 - 9

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

Table 5-8 Instance #3 (Bit-Strobe)

#

Name

Access

Type

Comments

1

State

Get

USINT

0 = Nonexistent
1 = Configuring
2 = Waiting for Connections ID
3 = Established
4 = Timeout

2

Instance Type

Get

USINT

0001h (I/O Connection)

3

Transport Class Trigger

Get

BYTE

82h (Transport class and Trigger Server,
Cyclic, Class 2)

4

Produced Connection ID

Get

UINT

-

5

Consumed Connection ID

Get

UINT

-

6

Initial Comm Characteristics

Get

BYTE

Produces over message, group 1
Consumes over message, group 2

7

Produced Connection Size

Get

UINT

0...8 bytes depending in Fieldbus Input
size

8

Consumed Connection Size

Get

UINT

0008h

9

Expected Packet Rate

Get/Set(a)

UINT

-

12

Watchdog Timeout Action

Get/Set(a)

USINT

0000h (Transition to the timed out state)

13

Produced Connection Path Length

Get

UINT

0007h

14

Produced Connection Path

Get

EPATH

20 04 25 64 00 30 03h

15

Consumed Connection Path
Length

Get

UINT

0007h

16

Consumed Connection Path

Get

EPATH

20 04 25 96 00 30 03h

17

Production Inhibit Time

Get

UINT

0000h

Advertising