6 connection object (05h) – INFICON Composer Elite Gas Concentration Monitor User Manual

Page 73

Advertising
background image

5 - 7

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

5.4.1.6 Connection Object (05h)

Connection objects are shown in

Table 5-6

.

The supported services are:

Class - Get Attribute Single (Revision, UINT, 0001h)

Instance - Get/Set Attribute Single

Table 5-6 Instance #1,#10...#14 (Explicit Messaging)

#

Name

Access

Type

Comments

1

State

Get

USINT

0 = Nonexistent
1 = Configuring
2 = Waiting for Connections ID
3 = Established
4 = Timeout
5 = Deferred Delete

2

Instance Type

Get

USINT

0000h (explicit messaging type)

3

Transport Class Trigger

Get

BYTE

83h (Server, Transport class 3)

4

Produced Connection ID

Get

UINT

-

5

Consumed Connection ID

Get

UINT

-

6

Initial Comm Characteristics

Get

BYTE

21 = Instance#1
33 = Instance #10..#14

7

Produced Connection Size

Get

UINT

262

8

Consumed Connection Size

Get

UINT

262

9

Expected Packet Rate

Get/Set(a)

UINT

2500ms

12

Watchdog Timeout Action

Get/Set(a)

USINT

0001h = Auto Delete (default)
0003h = Deferred delete

13

Produced Connection Path Length

Get

UINT

0000h (No connection path)

14

Produced Connection Path

Get

EPATH

-

15

Consumed Connection Path
Length

Get

UINT

0000h (no connection path)

16

Consumed Connection Path

Get

EPATH

-

17

Production Inhibit Time

Get

UINT

0000h

(a) Not able to be set on UCMM explicit message connections (Instance 10...14)

Advertising