4 sensors 1,2,3 error/warning data set – INFICON Composer Elite Gas Concentration Monitor User Manual

Page 84

Advertising
background image

5 - 18

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

5.4.2.2.4 Sensors 1,2,3 Error/Warning Data Set

This data set returns a compound set of current data values for three of the five
sensor cards.

Table 5-15 Sensors 1,2,3 error/warning data set

Name

Range

Data Type Note

Cached Data Set

7

Byte

Numeric value of the data set cached.

Boards Present
Indicator

0 – 0x1F

Byte

Bitwise: Bits 0 – 4 set indicate sensor board
1 – 5 present, respectively.

Sensor 1
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA

Refer to

Table 5-18

Integer

Status bits Refer to

Table 5-18

SS bit is not relevant in the data set format.

Sensor 1
Errors

------

Integer

Current error value Refer to

Table 5-22

for a

complete description of the errors.

Sensor 1
Warnings

------

Integer

Current warning value Refer to

Table 5-23

for

a complete description of the warnings.

Sensor 1
Sample Number

0 – 255

Byte

The current sample number.

Sensor 1
Frequency

Float

Sensor 1
Amplitude

Float

Sensor 1
Heater 1 Current
Drawn

Float

Sensor 1
Heater 2 Current
Drawn

Float

Sensor 2
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA

Refer to

Table 5-18

Integer

Status bits Refer to

Table 5-18

.

SS bit is not relevant in the data set format.

Sensor 2
Errors

------

Integer

Current error value. Refer to

Table 5-22

for a

complete description of the errors.

Sensor 2
Warnings

------

Integer

Current warning value Refer to

Table 5-23

for

a complete description of the warnings.

Sensor 2
Sample Number

0 – 255

Byte

The current sample number.

Sensor 2
Frequency

Float

Sensor 2
Amplitude

Float

Sensor 2
Heater 1 Current
Drawn

Float

Sensor 2
Heater 2 Current
Drawn

Float

Advertising