5 - 25

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

UZ

Zero Reference (Bit 28)

0|1

Bit

When sensor number is 1-5:
0= Sensor is using Factory Zero.
1= Sensor is using User Zero.

When Sensor Number is 0:
0= All the sensors are using the Factory
Zero.
1= At least one of the sensors is using the
User Zero.

AT

At Temperature (Bit 27)

0|1

Bit

When sensor number is 1-5:
0= Sensor is not at temperature (neither
heater is at temperature).
1= Sensor is at temperature (heater 1 within
+/-0.002ºC and heater 2 within +/-0.005°C).

When Sensor Number is 0:
0= Not all of the online sensors are at
temperature.
1= All the online sensors are at temperature.

CS

Concentration Steady
(Bit 26)

0|1

Bit

When sensor number is 1-5:
0= Sensor Concentration is currently not
steady.
1= Sensor Concentration is currently steady.

When Sensor Number is 0:
0= Not all the online sensors currently have
steady concentration.
1= All the online sensors currently have a
steady concentration.

Reserved (Bit 25)

ER

Errors (Bit 24)

0|1

Bit

When sensor number is 1-5:
0= Sensor currently does not have any
errors.
1= Sensor currently does have errors.

When Sensor Number is 0:
0= None of the sensors currently have
errors.
1= At least one of the sensors currently have
errors.

RS

Reset (Bit 23)

0|1

Bit

0= There has not been a power cycle/reset.
1= There has been a power cycle since the
previous command.

When sent sensor to controller:
bit indicates sensor card reset.

When sent control to external host:
bit indicates controller card reset or a box
power cycle. (bit cleared after returned).

Table 5-18 Status bit (continued)