4.1.10 What Happens if the Reference Zero

is Improperly Set?

If the Reference Zero is set when the conditions are wrong, i.e., not completely
purged of precursor gas:

1

A permanent concentration offset is introduced equal to the residual precursor
concentration when the new Reference Zero is applied. This is not corrected
until a proper Reference Zero is performed.

2

The sensitivity is altered, leading to an error proportional to the concentration.

Figure 4-6 Impact of wrong zero frequency

NOTE: Factory Zero will be restored if one of the carrier gas parameters

(molecular weight or

) is altered or if TEMP 1 Setpoint is changed or if

Allow User Zero is changed from Yes to No.