5.4.2.2.6 5 Sensors' Concentration Data Set

This data set returns a compound set of concentration values for all five of the
sensor cards.

Sensor 4

Heater 1 Current

Drawn

Float

Sensor 4

Heater 2 Current

Drawn

Float

Sensor 5

SS, LS, MH1/PC,

MH2/UZ, AT, CS,

ER, RS, BM, WA

Refer to

Table 5-18

Integer

Status bits Refer to

Table 5-18

.

SS bit is not relevant in the data set format.

Sensor 5

Errors

------

Integer

Current error value. Refer to

Table 5-22

for a

complete description of the errors.

Sensor 5

Warnings

------

Integer

Current warning value. Refer to

Table 5-23

for

a complete description of the warnings.

Sensor 5

Sample Number

0 – 255

Byte

The current sample number.

Sensor 5

Frequency

Float

Sensor 5

Amplitude

Float

Sensor 5

Heater 1 Current

Drawn

Float

Sensor 5

Heater 2 Current

Drawn

Float

Table 5-16 Sensors 4,5 error/warning data set

Name

Range

Data Type Note

Table 5-17 5 sensors' concentration data set

Name

Range

Data Type Note

Cached Data Set

9

Byte

Numeric value of the data set cached.

Boards Present
Indicator

0 – 0x1F

Byte

Bitwise: Bits 0 – 4 set indicate sensor board
1 – 5 present, respectively.

Sensor 1
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA

Refer to

Table 5-18

Integer

Status bits Refer to

Table 5-18

.

SS bit is not relevant in the data set format.

Sensor 1
Concentration

0.0 – 100.0

mole %

Float

The current concentration value.

Sensor 2
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA

Refer to

Table 5-18

Integer

Status bits Refer to

Table 5-18

.

SS bit is not relevant in the data set format.