5.6.3 Update Commands

Message = U <Command ID><Sensor><Internal Parameter><Data>

Command ID = <Byte> See Command ID in

Table 5-20

.

Sensor = <Byte> 1-5

Internal Parameter = <Byte> 0

Data = <Float> See range column in

Table 5-20

.

Response = None (header and trailer only)

Table 5-20 Query & Update commands

Q & U

Command

ID

Name Range

(low/high)

Data

Type

Note

Restrictions

2

Carrier Molecular Weight

1.000 -

1000.000

(AMU)

Float

This value represents the
molecular weight of the lighter gas.

-

3

Carrier Gamma

1.0000 -

2.0000

Float

This value represents the
Carrier’s specific heat ratio.

-

4

Precursor Molecular Weight

1.000 -

1000.000

(AMU)

Float

This value represents the
molecular weight of the
heavier gas.

-

5

Precursor Gamma

1.0000 -

2.0000

Float

This value represents the
Precursor’s specific heat ratio.

-

6

Allow User Zero

0-1

Integer Current value of Allow User Zero.

-

7

Reserved For Future Use

8

Reserved For Future Use

9

Query Zone 1 Temperature

30.00 -

65.00(ºC)

Float

Current value of the Zone 1
Temperature Setpoint parameter of
the Acoustic Sensor’s resonant
chamber.

-

Zone 2 Delta

-99.999 -

99.999(ºC)

Float

Current value of the Zone 2
Temperature Delta parameter of
the Acoustic Sensor’s resonant
chamber.

Will reject an

update for

Heater 2 if the

temperature is

not within the

30 to 68°C

range.

10

Result Averaging Filter Depth

0-100

Integer Current value of Averaging Filter

Depth.

-

20

Sets/Gets Bit Pattern for S0
Command

0 - 0x7FF8

0000

Encode Each bit represents a piece of data

returned. When a bit is set (1) that
piece of data is returned in the S0
command response. Refer to

Table

5-18

for details of which byte

relates to what data.
Bit = 0 Data not sent.
Bit = 1 Data is sent.