Table 5-22, Value refer to, Refer to – INFICON Composer Elite Gas Concentration Monitor User Manual

Page 98: Table 5-22 on

Advertising
background image

5 - 32

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

1

Current
Temperature

Temperature Zone 1|2

30.0 -

65.00(ºC)

99.999

Float

Current Zone Temperature 1 for
Acoustic Sensor's resonant chamber.
(99.999 if cable is disconnected.)

30.0 -

68.00(ºC)

99.999

Float

Current Zone Temperature 2 for
Acoustic Sensor's inlet tube. (99.999 if
cable is disconnected.)

3

Heater Status

Heater ID 1|2

0-6

Short

Status of Acoustic Sensor's resonant
chamber heater (1) and inlet tube
heater (2):
0 = Off
1 = Not at Temperature
2 = At Temperature
3 = Failed
4 = Cable Disconnected
5 = Near Temperature
6= Over Temperature

7

Boards
Present

Active Sensor Boards.

Bits 0-4 indicate sensor

board 1-5 present.

0x1F

Byte

0 = board not present.
1 = board present.

8

Keyboard Lock
State

0|1

Byte

0 = Locked, 1 =Unlocked.

9

Errors/
Warnings
Code

Errors

0 - 0xFFFF

Short

Current error value. See

Table 5-22

.

Warnings

0 - 0xFFFF

Short

Current warning value. See

Table

5-23

.

Table 5-22 System and hardware errors

Error

Number

Description

Error Code

Set by

CC1

or SC2

Sensor

Specific

(SS3)

Set/Clear

Trigger

General Notes/Actions

32

Manual Parameter
Reset Required

0x8000 0000

CC

Does not

clear

ACTION(CC): Parameters will
lock internally. Operation will be
suspended.
NOTE: Expected to occur
because of any errors that occur
even though the flash versions
are compatible.

31

Parameter Flash HW
Failure

0x4000 0000

CC

Does not

clear

ACTION(CC): Parameters will
lock internally. Operation will be
suspended.
NOTE: Indicates that the flash is
not responding correctly to the
read/write requests.

30

Image Flash HW
Failure

0x2000 0000

CC

Cleared on

read

NOTE: Indicates that the flash is
not responding correctly to the
read requests.

1

CC = Controller Card,

2

SC = Sensor Card,

3

SS = Specific Sensor

Table 5-21 Status commands (continued)

S

Command

ID

Name

Parameter

Range

(low/high)

Data

Type

Note

Advertising