Composer Elite Operating Manual

2.1.3 Composer Elite Location Selection

2.1.3.1 Selecting The Acoustic Sensor Location

Locate the Acoustic Sensor at a point after the Dilution Flow gas is added and
before the Back Pressure Controller. (See

Figure 2-2 on page 2-6

.) This creates

the environment with the most stable pressure and flow conditions for best
Composer Elite stability. Positioning the Acoustic Sensor after the Dilution Flow
also simplifies calibration by making it easy to flow pure carrier gas through the
Acoustic Sensor to set the Reference Zero.

Figure 2-2 Basic integration with a reactor

The Acoustic Sensor may be installed in any orientation, however, INFICON
recommends the normal inlet and outlet tube convention (refer to

Figure 2-2

) be

followed. The inlet tube is substantially longer than the outlet tube to facilitate
temperature equalization. If the inlet and outlet tube connections are reversed,
there may be a small frequency or concentration offset when the total gas flow
through the sensor is varied.
Ventilation clearance is not required; however, the surface the Acoustic Sensor is
mounted to must be at least 5°C below Temperature Setpoint 1.
The dimension from inlet tube VCR-4 sealing surface to outlet tube VCR-4 sealing
surface is 124 mm (4.88 in.).
Acoustic Sensor installation or removal from process tubing requires side-to-side
or top-to-bottom clearance of 152.4 mm (6 in.) for wrench rotation.
Interconnect Cable installation or removal requires 76.2 mm (3 in.) clearance.

Mains

Power Optional

Power

Supply

24 V(dc)

RS-232

Sensor

Control

Unit

DeviceNet

Interconnect

Cable

Ca

rr

ie

r Ga

s

P

MFC

MFC

Bubbler

Dilution Flow

Inlet

Acoustic Sensor

Located downstream

Outlet

Back

Pressure

Controller

Reactor

MFC (Mass Flow Controller)

from Dilution Flow