2.1.2.3 Composer Elite System Grounding Requirement

CAUTION

The Sensor Control Unit and Acoustic Sensor must be
grounded for proper performance of Composer Elite.

The following grounding method is recommended:

1

Verify that the Reactor is connected to a good, reliable earth ground (refer to

section 2.1.2.2, Earth Ground Requirement, on page 2-3

). Use a solid copper

strap, at least 0.08 cm (0.030 in.) thick by 2.5 cm (1 in.) wide, and as short as
practical, for the connection between Reactor and earth ground (see

Figure 2-1

).

2

Use a solid copper strap, approximately 0.08 cm (0.030 in.) thick by
2.5 cm (1 in.) wide, and as short as practical, between the Sensor Control Unit
ground stud (see

Figure 2-1

) and the Reactor. A ring terminal soldered to the

ground strap will allow for a good connection to the ground stud and convenient
removal or installation.

3

Connect the Acoustic Sensor to earth ground by one or both of the
following methods:

3a

Use the M4 mounting holes provided on the Acoustic Sensor to attach the
Acoustic Sensor to a metal mounting surface that is connected to earth ground
(see

Figure 2-1

and

section 2.2.2 on page 2-9

).

or

3b

Connect the Acoustic Sensor's outlet tubing to process tubing that is connected
to earth ground (see

Figure 2-1

and

section 2.2.2 on page 2-9

).

Figure 2-1 System grounding diagram

Sensor Control Unit

Ground

Stud

Earth Ground

Acoustic Sensor

Reactor

Grounded

Copper Straps

Process

Tubing

Grounded

Mounting

Surface