2.2.2 Installing The Acoustic Sensor

WARNING

Many gases used for film growth are toxic at very low
exposure levels.

CAUTION

Although the Acoustic Sensor was thoroughly leak
tested at the factory, periodic leak testing of the Acoustic
Sensor is recommended. See

section 7.2, Leak Test

Procedures, on page 7-1

.

1

Remove the caps from the VCR-4 fittings on the Acoustic Sensor's inlet and
outlet tubes by holding the fitting stationary with a 5/8 in. wrench while turning
the cap counterclockwise with a 3/4 in. wrench. Do not discard the caps.

2

Remove and discard the gaskets from the VCR-4 fittings.

3

Mount the Acoustic Sensor securely to a rigid surface, using the four holes with
M4 thread, to prevent stress at the inlet and outlet tubing welds and seals. See

Figure 2-3

and

Figure C-2 on page C-3

.

CAUTION

Do not use the inlet and outlet tubing to support the
weight of the Acoustic Sensor.

CAUTION

Maximum allowable depth of screw penetration into the
M4 mounting holes is 6.4 mm (0.25 in.).

Figure 2-3 M4 mounting holes