INFICON Composer Elite Gas Concentration Monitor User Manual

Page 101

Advertising
background image

5 - 35

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

19

All CC/SC Command
Retries Have Failed

0x0004 0000

CC

SS

Cleared on

read

NOTE: CC has transmitted a
command to the SC the
maximum number of times
without receiving a response.
Attempt has been abandoned.

18

Sensor Offline

0x0002 0000

CC

SS

Self Clear

17

Sensor firmware
Changed

0x0001 0000

CC

SS

Cleared on

read

15

No Signal (Cable
detected)

0x0000 4000

SC

SS

Self Clear NOTE: No/very low (below

operating threshold) audio
signal. Sensor card will
continually be in
Search mode.

14

Concentration Result
Error

0x0000 2000

SC

SS

Self Clear NOTE: Generally indicates a

change in gas flow/mixture. SC
will perform new sweep if it
persists more than
a few seconds.

13

Bad peak Shape

0x0000 1000

SC

SS

Self Clear NOTE: Indicated by low

R2

5

/ high Std Err.

Table 5-23 System and hardware warnings (continued)

Warning

Number

Description

Warning Code Set by

CC

1

or SC

2

Sensor

Specific

(SS

3

)

Set/Clear

Trigger

General Notes/Actions

1

CC = Controller Card,

2

SC = Sensor Card,

3

SS = Specific Sensor,

4

ABIC = Anybus Integrated Circuit,

5

R2= Lorentzian Coefficient of Determination

Advertising