1 hello command, 2 query commands, Section 5.6.1 – INFICON Composer Elite Gas Concentration Monitor User Manual

Page 94

Advertising
background image

5 - 28

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

5.6.1 Hello Command

Message = H <Command ID> <Sensor> <Internal Parameter>

Command ID = <Byte> 0-1

Sensor = <Byte> 0-5 The control card is represented with a 0.

Internal Parameter = <Byte> 0

Response = <String>|<Byte> See

Table 5-19

.

5.6.2 Query Commands

Message = Q <Command ID><Sensor><Internal Parameter>

Command ID = <Byte> See Command ID in

Table 5-20

.

Sensor = <Byte>1-5

Internal Parameter = <Byte> 0

Response = <Float> See range column in

Table 5-20

.

Table 5-19 Hello command response

H

Command

ID

Name

Value

Range

(low/high)

Data

Type

Note

0

Hello String

Composer Elite ver. xx.xx.xx

8x21

String The model and version numbers

of Composer Elite,
'Composer Elite ver xx.xx.xx'
where xx.xx.xx is the version
number.

1

Hello Binary

Version Major

0-99

Byte

Major Version Number.

Version Minor

0-99

Byte

Minor Version Number.

Version Build

0-99

Byte

Build Version Number.

Advertising