1 hello command, 2 query commands, Section 5.6.1 – INFICON Composer Elite Gas Concentration Monitor User Manual
Page 94

5 - 28
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
5.6.1 Hello Command
Message = H <Command ID> <Sensor> <Internal Parameter>
Command ID = <Byte> 0-1
Sensor = <Byte> 0-5 The control card is represented with a 0.
Internal Parameter = <Byte> 0
Response = <String>|<Byte> See
.
5.6.2 Query Commands
Message = Q <Command ID><Sensor><Internal Parameter>
Command ID = <Byte> See Command ID in
.
Sensor = <Byte>1-5
Internal Parameter = <Byte> 0
Response = <Float> See range column in
.
Table 5-19 Hello command response
H
Command
ID
Name
Value
Range
(low/high)
Data
Type
Note
0
Hello String
Composer Elite ver. xx.xx.xx
8x21
String The model and version numbers
of Composer Elite,
'Composer Elite ver xx.xx.xx'
where xx.xx.xx is the version
number.
1
Hello Binary
Version Major
0-99
Byte
Major Version Number.
Version Minor
0-99
Byte
Minor Version Number.
Version Build
0-99
Byte
Build Version Number.