3.1.1 Technical data for the VT1200 S

General data:

Sectioning frequency (± 10 %): ...................................................................................................................85 Hz (± 10 %)

Amplitude:

.......................................................................................................from 0 - 3 mm, in increments of 0.05 mm

Sectioning speed (± 10 %): ........................................................................................................................0.01 - 1.5 mm/s

Return speed (± 10 %): ....................................................................................1.0 - 5 mm/s, in increments of 0.5 mm/s

Total vertical specimen stroke: ......................................................................................................... 20 mm (motorized)

Sectioning range: ..................................................................................................................................................... 45 mm

Sectioning window: .................................................................................................................................. 0.5 mm - 45 mm

Specimen retraction: ..............................................................................0 - 100 µm (adjustable; can be deactivated)

Maximum specimen size:

with standard blade holder: ..................................................................................................................... 33 x 50 mm

Specimen orientation, rotating: ...........................................................................................................................360°

Specimen plate, swiveling: ...............................................................................................................................0 - 10°

Sectioning thickness setting: ............................................manual in 1 µm increments or automatic max. 1000 µm

Ambient conditions:

Operating temperature range: .....................................................................................................min. 10 °C - max. 35 °C

Relative humidity: ............................................................................................................................................... max. 60 %

Storage temperature: .............................................................................................................................................5 - 55 °C

Storage humidity: .......................................................................................................................................................< 60 %

Height:

.................................................................................................................................up to 2000 m above sea level

Electrical data:

Rated voltage range (± 10 %): ...................................................................................................................... 100 V - 240 V

Nominal frequency (± 10 %): ................................................................................................................................ 50/60 Hz

Power consumption: ..................................................................................................................................................35 VA

Protective class: ..................................................................................................................................................................I

Power fuse: ..................................................................................................................................................... T 1 A L 250 V

Pollution degree: .................................................................................................................................................................2

Overvoltage category: ...................................................................................................................................................... II

Electrical overload protection: .................................................................................................................................... Yes

Internal current limit for the electronics: .................................................................................................................. Yes

Dimensions:

L x W x H: .............................................................................................................................. 600 mm x 250 mm x 230 mm

Height with magnifier: ......................................................................................................... 600 mm x 250 mm x 320 mm

Height with microscope: .................................................................................................... 600 mm x 250 mm x 490 mm

Control unit (when the bases are folded in): ..................................................................... 190 mm x 150 mm x 72 mm

Weight:

Without magnifier support and control unit: ...........................................................................................................56 kg

VT1200 S control unit: ...................................................................................................................................................1 kg

Magnifier support: .........................................................................................................................................................2 kg

Microscope support with stereomicroscope: ..........................................................................................................4 kg