26

Instructions for Use V 1.4 RevC – 09/2013

5.

Working with the Instrument

Key / setting dial

Semiautomatic sectioning mode Automatic sectioning mode
If the DOWN key is pressed a

second time while the specimen

receptacle is moving downwards,

the specimen receptacle comes

to a stop and the current posi-

tion appears on the ∑µm display

(lowest position = 0, uppermost

position = 20,000 µm). The ∑µm

display does not change while the

specimen receptacle is in motion.

ditto

When the UP key is pressed and

held down, the specimen recep-

tacle moves quickly upwards into

the desired position. After the UP

key has been released, the cur-

rent position of the specimen re-

ceptacle appears on the ∑µm dis-

play. If the specimen receptacle

reaches the upper end position,

the LED in the UP key lights up

(uppermost position = 20,000 µm).

The ∑µm display does not change

while the specimen receptacle is

in motion.

ditto

Blade forward

Blade back

The "Blade forward" and "Blade

back" keys have to be held down

until the desired position is

reached. The blade feed rate can

be set in the menu: 1 – 5 mm/s,

in increments of 0.5 mm/s. Each

time an end point is reached, the

corresponding LED lights up in

the key.

ditto