7. Malfunctions: Meaning and Troubleshooting

7.1 Error messages and troubleshooting

Error No. / INF

No.

Error

Troubleshooting

Comment

Error 01

- Wrong control panel

(VT1200 or VT1200 S)

- Use the correct control

panel for the instrument.

- Using an incorrect control

panel does not harm the

instrument, but no func-

tions whatsoever are

available.

Error 21

- Head does not swing into

place.

1. Check to see if an am-

plitude value has been

selected.

2. Try to make the head

swing into place by gently

striking it with your hand.

3. If head does not swing

into place, inform service.

Error 22

- Timeout while initializing

the x-axis. (Timeout)

- Inform service!

Error 23

- DC motor of x-axis does

not rotate (during initial-

ization or in normal op-

eration)

- Inform service!

Error 24

- "X-Start" limit switch is

not reached.

- Inform service!

Error 25

- "X-Start" limit switch can-

not be moved away from.

- Inform service!

Error 26

- "X-Stop" limit switch is not

reached.

- Inform service!

After all error messages, the instrument MUST be switched off, then on again using the main

switch.