25

Leica VT1200 / VT1200 S

5. Working with the Instrument

Key / setting dial

Semiautomatic sectioning mode Automatic sectioning mode

2

The rotary knob for moving the

specimen receptacle vertically

can be used for fine-adjusting

the blade-to-specimen proximity.

Turning the setting dial clockwise

moves the specimen up into the

desired proximity; turning the set-

ting dial counterclockwise lowers

the specimen (minus sign).

The step size: 1, 10 or 100 µm can

be selected by pressing the STEP

SIZE key.

After the setting dial is turned

clockwise or counterclockwise,

the specimen receptacle position

is updated in the ∑µm display.

In semiautomatic mode, the de-

sired section thickness is se-

lected using the setting dial.

The selected section thickness

appears in the µm display and the

current specimen receptacle po-

sition appears in the ∑µm display.

After each sectioning process

has been completed, the µm dis-

play is set to "0".

ditto

Not possible.

When the DOWN key is pressed,

the specimen receptacle auto-

matically moves quickly into the

lowest position. (The LED in the

DOWN key lights up when the

lower end position is reached.)

The ∑µm display is set to "0".

ditto