Operating the vibrocheck – Leica Biosystems VT1200/VT1200 S User Manual

Page 42

Advertising
background image

40

Instructions for Use V 1.4 RevC – 09/2013

The deviation of the height amplitude is shown

on the display in µm (1).

This value can be ei-

ther positive or negative.

A number (e.g. –0.3)

appears in the ∑µm display (2). This means a

counterclockwise rotation (because of the "-")

by a turn of 0.3 and reduces the height ampli-

tude to a minimum. (If there is no operation

sign, the rotation is in a clockwise direction

"+".)

If the value is "0", no improvement of the

height amplitude is possible.

Fig.

34

16

17

18

Fig. 33

7. If the measured value is accepted, press

DOWN (flashes green). VC moves to lowest

position - blade is moved to the rear. Control

panel of the VT shows: "VIBRO END". The LED

on the VC is again illuminated in red.

8. The instrument now expects the VibroCheck

to be removed. To do so, disconnect the USB

plug connection of the VC to the basic instru-

ment and pull the VC off of the dovetail guide.

LED in DOWN key continues to flash – wait

until LED in DOWN button goes out.

The normal operating status is now restored.

6.

Operating the VibroCheck

18a

18

2

1

5. Press the STOP key. Unscrew the clamping

screw (16) only slightly using the size 3 Allen

key, pull the cap (17) up and off (keep it in a

safe place) and turn the adjusting screw (18)

counterclockwise (the "-" direction, 18a) by a

turn of 0.3 using the size 3 Allen key. Tighten

the clamping screw (16) clockwise.

6. Press the RUN key, check the value, repeat

steps 5-7 if necessary.

Fig. 32

If the display shows µm (2) "0" (opti-

mal) and the value in µm (1) is unaccept-

ably high, the blade must be replaced.

Advertising