Malfunctions: meaning and troubleshooting – Leica Biosystems VT1200/VT1200 S User Manual

Page 46

Advertising
background image

44

Instructions for Use V 1.4 RevC – 09/2013

7.

Malfunctions: Meaning and Troubleshooting

Error No. /

INF No.

Error

Troubleshooting

Comment

Error 52

- Basic calibration of light

barrier is not possible for

VibroCheck.

- Emitter diode or receiver

is defective.

- VibroCheck is defective.

- Inform service!

Error 53

- VibroCheck blade search

not successful.

- No blade and/or blade

holder swung out or not

installed.

- Emitter diode or receiver

dirty. Clean

- VibroCheck is defective.

- VT1200 drives to the low-

est Z position using the

stepper motor after an

unsuccessful search.

(Allows blade installation)

Communication

between key-

board - VT1200/S

- Communication error

between control unit and

VT

- VT1200 S display lights

up, but remains blank.

- VT1200: row of dots flash-

es.

- Inform service!

- The knife/

blade collides

with the buffer

tray

- The buffer tray was not

pushed in up to the rear

pin or was not clamped

down during installation.

Push in the buffer tray as

far as it will go and clamp

it down with the lever.

- The Vibro-

Check collides

with the blade

- The VibroCheck was not

pushed all the way in or

was not clamped down

during installation.

Push in the VibroCheck as

far as it will go and clamp

it down with the lever.

- Switching from SINGLE to

CONT is not possible.

- In MAN mode, only single cuts are possible.

- The cutting window can-

not be defined.

- In MAN mode, only single cuts are possible.

- If multiple cuts are required, switch to AUTO mode.