Working with the instrument – Leica Biosystems VT1200/VT1200 S User Manual

Page 32

Advertising
background image

30

Instructions for Use V 1.4 RevC – 09/2013

5.

Working with the Instrument

Key / setting dial

Semiautomatic sectioning mode Automatic sectioning mode

1

2

Rotary knob 2

Mode: Choice between AUTO

and MAN with rotary knob 1; for

semiautomatic sectioning mode,

MAN must be selected.

Rotary knob 2

Mode: Choice between AUTO and

MAN with rotary knob 1; for the

automatic sectioning mode, AUTO

must be selected.

1

2

Rotary knob 2

Stroke type (CUT): Only single

stroke (SINGLE) can be selected

using rotary knob 1. If continuous

stroke (CONT) is selected, an au-

dible warning signal sounds.

Rotary knob 2

Stroke type (CUT): Choice be-

tween single stroke (SINGLE) and

continuous stroke (CONT) with

rotary knob 1.

1

2

Rotary knob 2

Specimen retraction (RETRACT)

cannot be set.

Value cannot be

changed. NO automatic

retraction is possible in

semiautomatic section-

ing mode.

Rotary knob 2

The specimen retraction (RE-

TRACT) can be set from 0 to

100 µm in 10 µm increments using

rotary knob 1.

Rotary knob 2

The LED illumination can be ad-

justed to 5 different levels of

brightness using rotary knob 1.

ditto