Operating the vibrocheck, 1 using the vibrocheck with the vt1200, Using the vibrocheck with the vt1200 – Leica Biosystems VT1200/VT1200 S User Manual

Page 39

Advertising
background image

37

Leica VT1200 / VT1200 S

6.1 Using the VibroCheck with the VT1200

Prior to assembly, the dovetail guide (3) is

brought down to the lowest position using the

DOWN key!
1. The user installs the VibroCheck: Push the VC

along the dovetail guide (3) to behind the mark

on the baseplate of the instrument (rear stop)

and clamp it down using the lever (15). Insert

the blade and clamp tightly. Return the blade

to the cutting position (

see Fig. 20

).

2. Plug the connecting plug of the VibroCheck

(VC) into the socket (2 ) on the left side panel.

Short red flash of LED on VC --> control panel

acknowledges VC. LED then remains red. The

LED in the DOWN key flashes green.

(2) - VibroCheck

Fig. 23

Control panel for VT

Foot switch for VT

3

3. The user presses the DOWN key. The VC

moves to the bottom position, after that the

blade moves to the rearmost position—the

LED in the RUN/STOP key flashes. Short red

flash of LED on VC --> it then remains red.

4. The user presses the RUN/STOP key: First, the

blade moves forward (into the position exactly

above the VibroCheck), then the VC moves into

a position in which the blade partially covers

the light barrier. (If the VC does not receive

any signal via the light barrier, the operation

is canceled and DOWN is enabled.)

LED on VC flashes green - RUN/STOP flashes

yellow.

We recommend using the VibroCheck after each blade change to check the

optimum position of the blade and minimize the vertical vibration.

6. Operating the VibroCheck

15

Fig. 24

The following instructions must be adhered to exactly. Noncompliance can cause

serious damage to the instrument.

Fig. 25

Advertising