3 moving the blade, Working with the instrument – Leica Biosystems VT1200/VT1200 S User Manual

Page 22

Advertising
background image

20

Instructions for Use V 1.4 RevC – 09/2013

5.

Working with the Instrument

5.2.2 Fine-adjusting the blade-to-specimen proximity and feeding to the section thickness

The rotary knob 2 for moving the specimen receptacle vertically can be

used both for fine-adjusting the blade-to-specimen proximity and for adjust-

ing the desired section thicknesses. The step size: 1, 10 or 100 µm can be

selected by pressing the STEP SIZE key.

Turning the setting dial clockwise carries out the desired feed; turning the

setting dial counterclockwise lowers the specimen (minus sign). The value

appears in the lower display when "µm" is selected. After each sectioning

process has been completed, the µm display is set to zero.

Save a commonly used section thickness by adjusting the setting dial for

the section thickness and pressing the MEMO key for 3 seconds. An audible

signal confirms that the value has been accepted.

Feed to the saved value by quickly pressing the MEMO key. Pressing the

MEMO multiple times feeds to the value multiple times.

Blade

forward back

Neither negative values nor a section thickness of over 1000 µm are

permitted. If an attempt is made to save impermissible values, an au-

dible warning signal sounds three times and the last (permissible)

value is retained.

5.2.3 Moving the blade

The "Blade forward" and "Blade back" keys have to be held down until the

desired position is reached. The blade feed rate is 2.5 mm/s. When each end

point has been reached, the corresponding LED lights up in the key.