Leica VT1200 / VT1200 S

6. Operating the VibroCheck

Prior to assembly, the dovetail guide (3) is

brought down to the lowest position using the

DOWN key!
1. The user installs the VibroCheck: Push the VC

along the dovetail guide (3) to behind the mark

on the baseplate of the instrument (rear stop)

and clamp it down using the lever (15). Insert

the blade and clamp tightly. Return the blade

to the cutting position (

see Fig. 19

).

2. Plug the connecting plug of the VibroCheck

(VC) into the corresponding socket on the left

side panel. Short red flash of LED on VC --> it

then remains red. Control panel detects VC

(

see Fig. 30

). The LED in the DOWN key flashes

green.

3. The user presses the DOWN key. The VC

moves to the bottom position, in addition to that

the blade moves to the rearmost position—the

LED in the RUN/STOP key flashes. Short red

flash of LED on VC --> it then remains red.

4. The user presses the RUN/STOP key: First, the

blade moves forward (into the position exactly

above the VibroCheck), then the VC moves

into a position in which the blade partially

covers the light barrier. LED on VC flashes

green - RUN/STOP flashes yellow. Control

panel shows: "VIBRO search" (

see Fig. 30

). A

search can last up to 1 minute. When LED on

VC is green and LED in RUN/STOP is yellow,

the blade starts to vibrate.

VibroCheck

Fig. 28

Control panel for VT

Foot switch for VT

6.2 Using the VibroCheck with the VT1200 S

Fig. 31

15

3

Fig. 30

We recommend using the VibroCheck after each blade change for checking the optimum posi-

tion of the blade and important sectioning parameters.

Fig. 29

The following instructions must be adhered to exactly. Noncompliance can cause seri-

ous damage to the instrument.

Advertising