5 inserting the blade, 6 adjusting the clearance angle, Working with the instrument – Leica Biosystems VT1200/VT1200 S User Manual

Page 37

Advertising
background image

35

Leica VT1200 / VT1200 S

5. Working with the Instrument

5.4.5 Inserting the blade

• Insert a size 3 Allen key sideways into the

blade holder through the hole (8) and rotate it

90° clockwise.

Clamp the blade as follows:

• Insert the provided size 3 Allen key from the top

through the opening (10) into the blade holder

(11) and open the blade holder (BH).

• Hold the entire razor blade (9) (not separated)

on the left and right with both hands and insert

it into the blade holder. Hook in the blade over

bottom pressure plate (

see Fig. 21

).

• Clamp down the BH by turning the size 3 Allen

key clockwise until hand-tight.

8

5.4.6 Adjusting the clearance angle

Fig. 20

12

13

14

Fig. 22

Fig. 21

10

11

9

Now return the blade holder to the cutting posi-

tion.
• To do so, insert a size 3 Allen key sideways

into the blade holder through the hole (8) and

rotate approx. 90° counterclockwise.

For more information about adjusting the clear-

ance angle, see

Fig. 22

.

• Insert a size 3 Allen key sideways into the

blade holder through the hole (8) and rotate it

to the desired clearance angle mark.

12 - 15°

13 - 18°

14 - 21°

The blade holder can be used for razor

blades, injector blades and sapphire

blades (universal blade holder).

The clamping screw (11) on the blade

holder must not be tightened too much!

For 15°, the effective clearance angle is "0".
The most commonly used setting is 18°

(13).