5.1 Description of the typical application

The Leica VT1200 / VT1200 S is a microtome with a vibrating blade and is

predominately used for sectioning fixed and unfixed specimens during

neurological research.

• To simplify the process of inserting the speci-

men, you can quickly move the specimen re-

ceptacle to the lowest position by pressing the

DOWN button.

• To prepare high-quality sections, especially

for unfixed tissue, Leica suggests using the

optional VibroCheck measuring instrument to

determine the height amplitude of the blade

after each blade change, and then to minimize

it using the setting screw on the blade holder.

To perform the aforementioned process, install

the VibroCheck instrument (

see p. 37 for the

VT1200 or p. 39 for the VT1200 S

), install the

blade and adjust the clearance angle to the

desired position. Then take the measurement

and make the appropriate adjustments to the

placement of the blade holder.

Remove the VibroCheck according to the in-

structions, and turn the blade 90° toward the

top to ensure that the ice tray and buffer tray

can be installed safely.

• Insert the buffer tray into the ice tray and cover

it with the lid. Fill the ice tray with crushed ice.

• Remove the cover and fill the buffer tray with

pre-cooled physiological buffer.

• Push the ice tray and buffer tray onto the

dovetail guide and clamp them down.

• Use cyanoacrylate adhesive to adhere the

specimen to the specimen plate and use a

manipulator to place it in the buffer tray. Insert

the hose for gassing the buffer into the hose

clamp.

• Use the UP button to lift the specimen to the

sectioning level of the blade quickly. Fine-

adjusting the blade-to-specimen proximity can

be performed using the desired step size 1, 10

or 100 µm.

• Advance the blade toward the specimen using

the "Blade forward" key.

• After sectioning the specimen, use the setting

dial to feed to the desired section thickness

using the selected step size (1, 10 or 100 µm).

This raises the specimen receptacle to the

desired value.

• Start the sectioning process by pressing the

RUN/STOP key. The process can be stopped

again by pressing this key after the cut has

been completed. Use the "Blade back" key to

move the blade in front of the specimen. Select

the section thickness for the next cut and start

the sectioning process again.

• After completing the sectioning process,

remove the blade, dispose of the specimen

receptacle by pushing the DOWN key into its

lowest position, and unclamp, empty and clean

out the ice tray and buffer tray.