Leica VT1200 / VT1200 S

7. Malfunctions: Meaning and Troubleshooting

Error No. /

INF No.

Error

Troubleshooting

Comment

- Not accepted:

• Negative values

• Values of more than

1000 µm

- In MAN mode, pressing

the AUTO FEED key feeds

the last saved value.

Blocked when

RUN is active!

When a key is pressed, a

beep sounds.

- In MAN mode, if the

PAUSE key has not been

assigned a function, a

short beep sounds.

If the 1000 hour limit is exceeded while the interment is in normal operating

mode, a flashing "SEr" appears in the upper 3-digit LED for approx. three sec-

onds after the sectioning process has been ended with the RUN/STOP key.

This indicates that the vibrating component needs to be serviced.

The instrument

is not working.

The plug connections are

loose or the instrument is

connected to the power

supply improperly or not at

all.

The power fuses are defec-

tive.

VT connection cable: Check

the control unit and power

plug.

Replace the power fuses.