Leica Biosystems VT1200/VT1200 S User Manual

Page 26

Advertising
background image

24

Instructions for Use V 1.4 RevC – 09/2013

5.

Working with the Instrument

Key / setting dial

Semiautomatic sectioning mode Automatic sectioning mode

1

The blade feed rate can be ad-

justed from 0.01 to

1.5 mm/s:

0.01 - 0.1 in 0.01 mm/s increments,

0.10 - 0.5 in 0.02 mm/s increments,

0.50 - 1.5 in 0.10 mm/s increments.

ditto

1

Adjusting the amplitude from

0 to 3 mm in 0.05 mm increments

ditto

1

Not possible.

Setting the section thickness for

automatic mode - max. 1000 µm.

Display for the current speci-

men holder position (lowest posi-

tion = 0 µm, uppermost position =

20,000 µm.)

You can reset the display to

"0" at any point by pressing the

CLEAR/∑µm key. The section

thicknesses are then added in the

µm display.

ditto