Leica VT1200 / VT1200 S

5. Working with the Instrument

5.4.3 Preparing a specimen

• Fix the specimen to the specimen plate using

the cyanoacrylate (included in the standard

scope of delivery).

• Screw the manipulator (6) onto the specimen

plate, place it in the buffer tray and adjust it to

the desired position.

• The specimen plate is held in place in the buf-

fer tray magnetically.

Fig. 18

6

5.4.2 Installing the double-walled buffer tray

Clamps for holding the gassing hose for the buf-

fer in the proper position can be added to the

double-walled buffer tray.

Fig. 17

Connect hoses (2, in the scope of delivery of the

double-walled buffer tray) to the empty buffer

tray (bottom of Fig. 17). Access is easier if you

make the left connection first. To do this, pull

back the closure coupling (11), fit the hose until

it engages audibly, and then make the right-hand

connection.

11

11

2

When using the double-walled buffer

tray, the recirculating cooler/chiller

must be installed PRIOR TO working

with specimens.