5.3 Control elements on the VT1200 S control panel

The Leica VT1200 S is a fully automatic microtome with a vibrating blade

that can be operated in semiautomatic or automatic sectioning mode.

5.

Working with the Instrument

Key / setting dial

Semiautomatic sectioning mode Automatic sectioning mode

Switch on the instrument.

1 = on 0 = off

When the instrument is switched

on, to make it easier to insert a

specimen, the specimen recep-

tacle automatically moves into

the lowest position (the LED in the

DOWN key lights up) and the blade

holder moves into the rearmost

position (the LED in "Blade away"

from the specimen lights up).

If semiautomatic sectioning

mode was selected before the

instrument was last switched

off, the following saved param-

eters are called up again once it

is switched back on:

• Blade feed rate (SPEED),

• Selected amplitude (AMPL)

ditto

• Blade feed rate (SPEED),

• Selected amplitude (AMPL)

• Saved section thickness (AUTO

FEED)

The LED illumination is switched

on automatically. You can switch

it off using the ON/OFF key.

ditto

The LED MAN is active.

semiautomatic sectioning mode

is enabled.

The LED AUTO is active.

The automatic sectioning mode

is enabled.