Malfunctions: meaning and troubleshooting – Leica Biosystems VT1200/VT1200 S User Manual

Page 44

Advertising
background image

42

Instructions for Use V 1.4 RevC – 09/2013

7.

Malfunctions: Meaning and Troubleshooting

Error No. /

INF No.

Error

Troubleshooting

Comment

Error 27

- "Bottom" limit switch of

the Z-axis is not reached

during initialization or

operation.

- Check whether an ob-

stacle is blocking the

travel path of the ice tray.

Remove the obstacle. If

there is no obstacle and

the error message per-

sists after the instrument

is switched on again:

Inform service.

Error 28

- The "top" limit switch of

the Z-axis is not reached.

- Check whether an ob-

stacle is blocking the

travel path of the ice tray.

Remove the obstacle. If

there is no obstacle and

the error message per-

sists after the instrument

is switched on again:

Inform service.

Error 31

- Both X sensors activated

(during initialization or in

normal operation)

- Inform service!

Error 32

- Both Z sensors activated

(during initialization or in

normal operation)

- Inform service!

After all error messages, the instrument MUST be switched off, then on again using the main

switch.