Leica VT1200 / VT1200 S

5. Working with the Instrument

Key / setting dial

Semiautomatic sectioning mode Automatic sectioning mode
In semiautomatic sectioning

mode, before each cut, a manual

feed to the desired section thick-

ness must be carried out using

the setting dial for section thick-

ness.

There is no automatic specimen

retraction in this mode; howev-

er, retraction can be performed

manually.

In automatic mode, the selected

section thickness feed (AUTO

FEED) is carried out along the 1st

selected cutting window edge

automatically before each cut. To

prevent the specimen surface and

the blade from coming into con-

tact while the blade is being re-

tracted, the specimen is lowered

by the desired retraction value

along the 2nd cutting window

edge after each complete cut.

Toggling from AUTO to

MAN

Toggling from MAN to

AUTO

The following key functions are

disabled in semiautomatic mode:

• Setting cutting window edges

• Selecting the continuous

stroke (CONT)

• Selecting the section thickness

for the automatic feed (AUTO

FEED)

PAUSE

not active.

An audible warning signal sounds

when these keys are pressed.

The following functions keys be-

come enabled again in automatic

mode:

• Cutting window edges that

have already been set

• Section thickness (AUTO FEED)

and continuous stroke (CONT)