3 safety features, 1 locking the handwheel, 2 knife guard – Leica Biosystems CM1510-S User Manual

Page 11: Safety features, Locking the handwheel, Knife guard, Safety

Advertising
background image

11

Leica CM1510 S – Cryostat

2.3

Safety features

The instrument is equipped with the following safety features: hand-
wheel locking mechanism and knife guard on knife holder.
The consistent use of these safety features and strict observation of
the safety instructions and warnings in this manual will safeguard the
operator from accidents and/or personal injury to a great extent.

2.3.1 Locking the handwheel

To lock the handwheel, rotate the handwheel until the handle is in the
upper position and the metal pin (1) is located opposite the black dot (2).
Push the metal pin outward into the recess at the handwheel; if neces-
sary move the handwheel slightly back and forth, until you feel that the
locking mechanism engages.

To unlock the handwheel, push the metal pin (1) out of the recess at the
handwheel, towards the cryostat housing.

Prior to manipulating knife and specimen, when changing the
specimen or during breaks, always lock the handwheel and
cover the cutting edge with the knife guard!

2.3.2 Knife guard

Knife holder CN is equipped with a knife guard; on the knifeholder CE, the
glass plate of the anti-roll guide constitutes the knife guard (see sepa-
rate instruction manual for your knife holder).

2.

Safety

2

1

Fig. 2