44

Instruction manual V 1.4 – 04/2010

5

•

Lift microtome slightly and pull it to the left to
disengage the plastic coupling (5) connect-
ing the axles.

•

Take microtome out of cryochamber.

Fig. 28

8.5

Removing the microtome housing

The microtome cover may be re-
moved to expedite thorough drying
of the microtome in an oven.
Place the microtome in an oven at 40
°C to 50 °C for several hours.
After repeatedly drying the micro-
tome in this manner, it is recom-
mended to relubricate the cross
roller bearings inside the microtome!
For further information, please con-
tact your Leica sales company!

•

Loosen the two screws (8) on both sides of
the housing.

•

To remove, pull the housing upwards. The
front plate of the microtome with the speci-
men head remains in place.

8

8

8.