5 removing the microtome housing, Removing the microtome housing, Cleaning and maintenance – Leica Biosystems CM1510-S User Manual

Page 44

Advertising
background image

44

Instruction manual V 1.4 – 04/2010

5

Lift microtome slightly and pull it to the left to
disengage the plastic coupling (5) connect-
ing the axles.

Take microtome out of cryochamber.

Fig. 28

8.5

Removing the microtome housing

The microtome cover may be re-
moved to expedite thorough drying
of the microtome in an oven.
Place the microtome in an oven at 40
°C to 50 °C for several hours.
After repeatedly drying the micro-
tome in this manner, it is recom-
mended to relubricate the cross
roller bearings inside the microtome!
For further information, please con-
tact your Leica sales company!

Loosen the two screws (8) on both sides of
the housing.

To remove, pull the housing upwards. The
front plate of the microtome with the speci-
men head remains in place.

8

8

8.

Cleaning and maintenance

Fig. 29