4.3

Unpacking and installation

Unpacking as per the unpacking instructions attached to the outside of
the instrument shipping crate.

The instrument must be transported either in an upright position
or slightly inclined, at an angle of max. 30 degrees!
When pushing the instrument, do not grip the cabinet at the lid.

Grip the cabinet only at the marked areas (

)!

Use the size 16 fork wrench (1) to com-
pletely screw in the adjustable feet.

Move instrument to installation site on roll-
ers (2 and 3).

The adjustable feet (1) can be used to
slightly tilt the instrument (max. 30 degrees).

To lift the instrument please use the handle
straps provided attaching them to the front
(2) and rear rollers (3).

At the installation site, unscrew the adjust-
able feet (1) with the fork wrench to ensure
stability.

Align the adjustable feet to level the instru-
ment.

After transporting, wait at least 4
hours before switching the instru-
ment on!
This waiting period is necessary to
allow the compressor oil, which may
have been displaced during trans-
port, to return to its original position.

Instrument
rear wall

2

3

1

4.

Setup

Fig. 5