6 reinstalling the microtome, Reinstalling the microtome, Cleaning and maintenance – Leica Biosystems CM1510-S User Manual
Page 45
45
Leica CM1510 S – Cryostat
8.6
Reinstalling the microtome
The microtome must be completely
dry before reinstallation.
Humidity inside the microtome will
condense and freeze in the cold cry-
ostat and thus may cause malfunc-
tions or damage.
•
Place the microtome into the cryochamber
slightly left from the original position. Make
sure the specimen head is in the lowest
position.
•
Lubricate the surface of the plastic cou-
pling (5) with a drop of cryostat oil.
•
Mount plastic coupling (5) on shaft (4).
•
Use your right hand to bring the handwheel
handle in the lowest position and keep the
handle in place.
•
Push the microtome to the right with your
left hand and, if necessary for proper align-
ment of the parts, slightly turn the hand-
wheel back and forth until coupling (5) en-
gages into shaft (6).
•
Tighten screws (3).
•
Reconnect the coarse feed motor plug (1)
and the temperature sensor (2) (ensure that
temperature sensor (2) fits against the cop-
per spring with the polished bush).
•
Put shelves, heat extractor, section waste
tray, brush shelf and knife holder back into
cryochamber.
•
Reinstall sliding window.
•
The instrument is ready to be switched
back on.
4
1
2
3
5
6
8.
Cleaning and maintenance
Fig. 30