Leica CM1510 S – Cryostat

8.6

Reinstalling the microtome

The microtome must be completely
dry before reinstallation.
Humidity inside the microtome will
condense and freeze in the cold cry-
ostat and thus may cause malfunc-
tions or damage.

•

Place the microtome into the cryochamber
slightly left from the original position. Make
sure the specimen head is in the lowest
position.

•

Lubricate the surface of the plastic cou-
pling (5) with a drop of cryostat oil.

•

Mount plastic coupling (5) on shaft (4).

•

Use your right hand to bring the handwheel
handle in the lowest position and keep the
handle in place.

•

Push the microtome to the right with your
left hand and, if necessary for proper align-
ment of the parts, slightly turn the hand-
wheel back and forth until coupling (5) en-
gages into shaft (6).

•

Tighten screws (3).

•

Reconnect the coarse feed motor plug (1)
and the temperature sensor (2) (ensure that
temperature sensor (2) fits against the cop-
per spring with the polished bush).

•

Put shelves, heat extractor, section waste
tray, brush shelf and knife holder back into
cryochamber.

•

Reinstall sliding window.

•

The instrument is ready to be switched
back on.

4

1

2

3

5

6

8.