Leica CM1510 S – Cryostat

Real time:

5.

Operation

Fig. 18

Fig. 17

Code parameter:

If ‘P‘ is pressed once more, the display
reads ‘C00‘. In this mode a code needs to
be entered.

Code parameter settings may be ac-
cessed only by authorized technical
service personnel.

Press ‘P‘ again to return set temperature indi-
cation, or wait 30 seconds until the display in-
dication returns to actual temperature.

Fig. 19

To set real time, press

once more.

Press

or

to modify the indicated

value (hour).

Press

again.

Press

or

to modify the indicated

value (minutes).

Minimum switch-off period

When the instrument is switched on (via mains plug), the compressor
starts operating after approx. 5 secs. If the compressor does not switch
on after that period of time, the

LED will start blinking, indicating that

the period during which the instrument was switched off was too short.
(Safety meaure to avoid major instrument damage).