Optional accessories, 1 mobile heat extractor, 2 stationary heat extractor – Leica Biosystems CM1510-S User Manual

Page 48: Mobile heat extractor, Stationary heat extractor

Advertising
background image

48

Instruction manual V 1.4 – 04/2010

•

Screw heat extractor into quick-freeze
shelf.

•

Place it directly on the specimen surface.
After approx. 30 seconds (depending on
tissue type and specimen size) the speci-
men is entirely frozen.

To prevent unnecessary distortion of
the specimen during contact freez-
ing, the weight of the heat extractor
can be balanced.

•

For that purpose, loosen screw (2) and ad-
just knob (1) upwards or downward.

•

Retighten screw (2) to fix knob (1) in the ap-
propriate position.

9.

Optional accessories

9.1

Heat extractors

Specimen freezing with the freezing shelf can be accelerated by the
additional use of a heat extractor.

9.1.1 Mobile heat extractor

•

Store the heat extractor in the cryochamber.

•

Place it on the specimen surface to accelerate freezing.

•

Remove the heat extractor once the specimen is entirely frozen.

Fig. 35

9.1.2 Stationary heat extractor

1

2

Fig. 36

Advertising