Troubleshooting, 1 possible causes and remedies, Possible causes and remedies – Leica Biosystems CM1510-S User Manual

Page 34

Advertising
background image

34

Instruction manual V 1.4 – 04/2010

7.

Troubleshooting

7.1

Possible causes and remedies

Frost on chamber walls and
microtome

Ice formation on the bottom of
the cryochamber.

Sections smear

Sections splinter

Sections not properly flattened

Problem

Cause

Remedy

- Cryostat is exposed to air cur-

rents (open windows and
doors, air conditioning).

- Sliding window open over a

longer period of time at a very
low chamber temperature.

- Frost built up by breathing into

the cryochamber.

- Drain of the quick freeze shelf

defrosting water obstructed.

- Specimen not cold enough.
- Knife/blade and/or anti-roll

plate not yet cold enough and
thus warm the sections.

- Specimen too cold

- Static electricity/air currents.
- Specimen not cold enough.
- Large area specimen.

- Anti-roll plate poorly posi-

tioned.

- Anti-roll plate poorly aligned

with knife edge.

- Incorrect clearance angle.
- Knife/blade blunt or damaged.

- Change place of installation

for the cryostat.

- Close sliding window correct-

ly.

- Wear mask if necessary.

- Switch off the instrument and

let it thaw and dry completely.

- Align the instrument with a

spirit level.

- Select lower temperature.
- Wait until knife/blade and/or

anti-roll plate have reached
cryochamber temperature.

- Select higher temperature.

- Remove cause.
- Select lower temperature.
- Trim the specimen parallel,

increase section thickness.

- Reposition anti-roll plate.

- Align correctly.

- Set correct angle.
- Use different part of the cut-

ting edge or replace.