Cleaning and maintenance, 1 general maintenance instructions, General maintenance instructions – Leica Biosystems CM1510-S User Manual

Page 39

Advertising
background image

39

Leica CM1510 S – Cryostat

8.1

General maintenance instructions

Only qualified and authorized service personnel may access the internal components of the
instrument for service and repair!

The microtome is virtually maintenance-free. Nevertheless, to ensure
smooth operation of the instrument over many years, we recommend
the following:

Have the instrument inspected by a qualified service engineer autho-
rized by Leica at least once a year.

Enter into a service contract at the end of the warranty period. For
further information, please contact your local Leica service center.

Clean the instrument every day.

From time to time - above all after the instrument has been dried re-
peatedly in a laboratory oven, apply a thin coat of cryostat oil onto:

-

clamping device (T-piece) and clamping levers of the microtome
base plate,

-

guiding device of the knife holder base,

-

clamping lever of the knife holder base,

- the specimen cylinder. To do so, move the specimen cylinder out

to the front limit stop (pressing the coarse feed button), apply a
few drops of cryostat oil onto the cylinder and once again push
the coarse feed button in opposite direction to move the speci-
men cylinder back to the rear limit position.

Visible dirt (dust fluff, etc.) on the ventilation grid of the condenser at
the bottom of the right cabinet side wall, should be removed with a
brush, broom or vacuum cleaner, cleaning in the direction of the fins.