Leica CM1510 S – Cryostat

6.4.1 Moving the specimen with the coarse / precision feed

The specimen movement towards the cutting edge must be ob-
served carefully to avoid that the specimen collides with the
cutting edge which can cause severe damage both to the knife
and the specimen!

Press the coarse feed button for fast or slow specimen movement
towards the cutting edge.

6.4.2 Section thickness setting

To select the section thickness, rotate knob (1) to a setting between 1
and 60 µm:
from 0 - 10 µm in 1-µm steps,
from 10 - 20 µm in 2-µm steps,
from 20 - 60 µm in 5-µm steps.

The graduation scale on the microtome reflects the selected section
thickness.

Select the desired section thickness via the rotary knob on the mi-
crotome.
Start sectioning at approx. 20 µm.

Decrease section thickness continually down to the appropriate value.
After changing from one section thickness to another, the first two or three sections should be
rejected.

When sectioning, turn the handwheel clockwise at a constant speed.