2 technical data, Technical data, Instrument components and specifications – Leica Biosystems CM1510-S User Manual

Page 14

Advertising
background image

14

Instruction manual V 1.4 – 04/2010

3.2

Technical data

Operating temperature range from +18 °C to +35 °C. All specifications related to temperature
are based on an ambient temperature of +22 °C and maximum air humidity of 60%!

Type

CM1510 S

CM1510 S

CM1510 S

Nominal voltage

100 V AC ±10%

120 V AC ±10%

230 V AC ±10%

Nominal frequency

50/60 Hz

60 Hz

50 Hz

Power draw

2000 VA

1300 VA

1500 VA

Protective class

I

I

I

Mains fuses

Automatic circuit

Automatic circuit

Automatic circuit

breaker T 20 A

breaker T 15 A

breaker T 10 A

Pollution degree

2

2

2

Overvoltage inst. category

II

II

II

Power draw

495 W / 850 W

850 W

495 W

Refrigerating output
at evaporator temperature

470 W / 780 W

780 W

470 W

of -25°C
Nominal pressure

25 bar (maximum value)

Safety factor

3

Refrigerant

215 g ± 5 g R404A *

Compressor oil

0.6 l EMKARATE RL244b, ICI *

Refrigeration system

50 Hz

60 Hz

Temperature setting range

0 °C to -30 °C ±10%
at an ambient temperature of 22 °C
and air humidity

_

<

60 %

when sliding window is closed;
adjustable in 1-degree increments

Temperature of quick-freezing shelf

Max. -44 °C
- at a temperature of -30 °C

Defrosting

Automatic hot gas defrosting,
defrost time freely programmable,
manual defrosting of quick-freeze shelf and evaporator
duration: 8 min.

* Refrigerant and compressor oil must be replaced by authorized service personnel only!

3.

Instrument components and specifications

Advertising