Computer – MTS Criterion 60 - Manual User Manual

Page 19

Advertising
background image

Description

MTS Criterion™ Series 60 Product Manual

Introduction

19

Mixing different brands of hydraulic fluid can affect the system
performance.

Contaminated hydraulic fluid can cause premature wear of the hydraulic
components in your system.

Do not mix different brands of hydraulic fluid. MTS System Corporation
recommends using Exxon Mobile DTE-25.

Computer

The computer is also an integral part of the system. It runs the testing software
which provides full machine control, data acquisition and management, and
advanced data analysis and presentation. MTS has minimized the amount of
custom electronics required for your system, thereby making it flexible and
reliable. This is done by connecting the frame and the computer via standard
USB 2.0 connectors. See the computer configuration below:.

Integrated Operations Platform Properties

P

ARAMETERS

64.305/64.305E

64.605/64.605E/

64.106/64.106E

Pump motor rating

1.5 kW (2.0 hp)

2.2 kW (2.9 hp)

Maximum
continuous pressure

23 MPa (3300 psi)

23 MPa (3300 psi)

Maximum flow
capacity (50 Hz)

3.3 L/min (0.9 gpm)

5.1 L/min (1.3 gpm)

Maximum flow
capacity (60 Hz)

4.0 L/min (1.0 gpm)

6.2 L/min (1.6 gpm)

Filtration

5 microns nominal

5 microns nominal

Reservoir capacity

85 L (22.5 gal)

85 L (22.5 gal)

CAUTION

Monitor and Computer Configuration

D

ESCRIPTION

Monitor

Model

L1710A

Brand

Lenovo Panel

Size

17 inch

Max Resolution

1280×1024

Description

Thinkvision LCD

Industrial PC

Model

ACP-4000

Advertising